Abstract

An improvement of a Hartshorn type alternating current magnetometer is reported. The use of a dual channel oscillator, a two-phase lock-in amplifier and a personal computer enabled us to obtain a simple feedback system for automatic balancing of the bridge. The resolution of this system is about 3×10-7emu of magnetic moment or higher at the field of 0.8 Oe in amplitude and 159 Hz in frequency.

Original languageEnglish
Pages (from-to)208-209
Number of pages2
JournalJapanese journal of applied physics
Volume29
Issue number1 R
DOIs
Publication statusPublished - Jan 1990
Externally publishedYes

Keywords

  • Ac susceptibility
  • Hartshorn bridge
  • High temperature superconductor
  • Magnetization measurement
  • Magnetometer
  • Spin glass