Abstract
In high-resolution biopotential measurements, interference from local 60 Hz power-line often disturbs the original signals so that it should be rejected as small as possible. The sources and elimination methods of this interference have been considered by many researchers. However, it does not completely eliminate because of many complex factors. We have designed a new method for eliminating the power-line interference by using an automatic interference controller device (AICD). The principle of this method is based on automatically matching of the amplitude and the phase from power-line signal detector of AICD on those of interference of biopotential signal. In this work, we used ECG signal as one of biopotential signals source. The device does not make ECG signal distorted and only eliminates interference signal. Interference elimination can be performed continuously and rapidly even if the situations of interference are changing with time or frequency in electrical system are varied. The handling is easy because of automatic system and it is applicable not only to ECG but also for the other biopotential signals.
Original language | English |
---|---|
Pages | 1358-1362 |
Number of pages | 5 |
Publication status | Published - Jan 1 2000 |
Event | IMTC/2000 - 17th IEEE Instrumentation and Measurement Technology Conference 'Smart Connectivity: Integrating Measurement and Control' - Baltimore, MD, USA Duration: May 1 2000 → May 4 2000 |
Other
Other | IMTC/2000 - 17th IEEE Instrumentation and Measurement Technology Conference 'Smart Connectivity: Integrating Measurement and Control' |
---|---|
City | Baltimore, MD, USA |
Period | 5/1/00 → 5/4/00 |
ASJC Scopus subject areas
- Electrical and Electronic Engineering