Abstract
Impedance techniques have been widely applied to biomedical engineering field. In order to obtain the definite results from bioimpedance, multi-channel measurement is effective. A linearity of biological tissue and fundamental technical parameters for the instrument must be confirmed. In this study, the fundamental conditions have been investigated for multi-channel bioimpedance measurement and the condition of measurement for biological tissue have been confirmed. The differences of every measurement frequency should be appropriate 1 kHz. And it has been cleared that the linearity of biological tissue is maintained. Based on these results, 2 channel bioimpedance measurement instrument has been constructed.
Original language | English |
---|---|
Pages | 231-234 |
Number of pages | 4 |
Publication status | Published - Jan 1 1998 |
Event | Proceedings of the 1998 IEEE Instrumentation and Measurement Technology Conference, IMTC. Part 1 (of 2) - St.Paul, MN, USA Duration: May 18 1998 → May 21 1998 |
Other
Other | Proceedings of the 1998 IEEE Instrumentation and Measurement Technology Conference, IMTC. Part 1 (of 2) |
---|---|
City | St.Paul, MN, USA |
Period | 5/18/98 → 5/21/98 |
ASJC Scopus subject areas
- Electrical and Electronic Engineering