T1 - Effects of exposure parameters and slice thickness on detecting clear and unclear mandibular canals using cone beam CT
AU - Jasa, Gainer R.
AU - Shimizu, Mayumi
AU - Okamura, Kazutoshi
AU - Tokumori, Kenji
AU - Takeshita, Yohei
AU - Weerawanich, Warangkana
AU - Yoshiura, Kazunori
N1 - Publisher Copyright:
© 2017 The Authors. Published by the British Institute of Radiology.
PY - 2017
Y1 - 2017
N2 - Objectives: The purpose of this study was to clarify the effects of exposure parameters and image-processing methods when using CBCT to detect clear and unclear mandibular canals (MCs). Methods: 24 dry half mandibles were divided into 2 groups with clear and unclearMCs based on a previous CBCT study. Mandibles were scanned using a CBCT system with varying exposure parameters (tube voltages 60 kV, 70 kV and 90 kV; and tube currents 2mA, 5mA, 10mA and 15mA) to obtain a total of 144 scans. The images were processed with different slice thicknesses using ImageJ software (National Institutes of Health, Bethesda, MD). Five radiologists evaluated the cross-sectional images of the first molar region to detect the MCs. The diagnostic accuracy of varying exposure parameters and image-processing conditions was compared with the area under the curve (Az) in receiver-operating characteristic analysis. Results: The Az values for clear MCs were higher than those for unclear MCs (p<0.0001). With increasing exposure voltages and currents, Az values increased, but no significant differences were found with high voltages and currents in clear MCs (p=1.0000 and p= 0.9340). The Az values of serial images were higher than those of overlaid images (p< 0.0001), and those for thicker slices were higher than those for thinner slices (p<0.0001). Conclusions: Our findings indicate that detection of unclear MCs requires either higher exposure parameters or processing of the images with thicker slices. To detect clear MCs, lower exposure parameters can be used.
AB - Objectives: The purpose of this study was to clarify the effects of exposure parameters and image-processing methods when using CBCT to detect clear and unclear mandibular canals (MCs). Methods: 24 dry half mandibles were divided into 2 groups with clear and unclearMCs based on a previous CBCT study. Mandibles were scanned using a CBCT system with varying exposure parameters (tube voltages 60 kV, 70 kV and 90 kV; and tube currents 2mA, 5mA, 10mA and 15mA) to obtain a total of 144 scans. The images were processed with different slice thicknesses using ImageJ software (National Institutes of Health, Bethesda, MD). Five radiologists evaluated the cross-sectional images of the first molar region to detect the MCs. The diagnostic accuracy of varying exposure parameters and image-processing conditions was compared with the area under the curve (Az) in receiver-operating characteristic analysis. Results: The Az values for clear MCs were higher than those for unclear MCs (p<0.0001). With increasing exposure voltages and currents, Az values increased, but no significant differences were found with high voltages and currents in clear MCs (p=1.0000 and p= 0.9340). The Az values of serial images were higher than those of overlaid images (p< 0.0001), and those for thicker slices were higher than those for thinner slices (p<0.0001). Conclusions: Our findings indicate that detection of unclear MCs requires either higher exposure parameters or processing of the images with thicker slices. To detect clear MCs, lower exposure parameters can be used.
KW - CBCT
KW - Computer-assisted image processing
KW - Image quality
KW - Mandible
UR - http://www.scopus.com/inward/record.url?scp=85019899660&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85019899660&partnerID=8YFLogxK
U2 - 10.1259/dmfr.20160315
DO - 10.1259/dmfr.20160315
M3 - Article
C2 - 28125294
AN - SCOPUS:85019899660
SN - 0250-832X
VL - 46
JO - Dentomaxillofacial Radiology
JF - Dentomaxillofacial Radiology
IS - 4
M1 - 20160315
ER -