T1 - Feasibility of 5-mm vs 2.5-mm width multileaf collimator in noncoplanar volumetric modulated arc stereotactic radiotherapy for multiple brain metastases
AU - Yoshio, Kotaro
AU - Wakita, Akihisa
AU - Hisazumi, Kento
AU - Kitayama, Takahiro
AU - Tajiri, Nobuhisa
AU - Shiode, Tsuyoki
AU - Akaki, Shiro
AU - Kanazawa, Susumu
N1 - Publisher Copyright:
© 2019 American Association of Medical Dosimetrists
PY - 2020/3/1
Y1 - 2020/3/1
N2 - The aim of this study was to examine the feasibility of noncoplanar volumetric modulated arc stereotactic radiotherapy (VMAT-SRT) using a 5-mm multileaf collimator (MLC) for multiple brain metastases. We identified 34 multiple-target cases (3 to 19 targets in each case) with a total of 257 of targets and constructed noncoplanar VMAT-SRT plans using 5-mm and 2.5-mm MLCs with 4-arc. The prescribed dose was 36 Gy/6 fr. Plans were evaluated using the Paddick conformity indices (PCI), Paddick gradient index (PGI), and normal brain dose (NBD, equal to the mean brain dose minus gross tumor volume). There were no significant differences in PCI (median [range]: 5 mm, 0.88 [0.78 to 0.94]; 2.5 mm, 0.89 [0.78 to 0.94]; p= 0.691), PGI (median [range]: 5 mm, 3.96 [2.21 to 6.63]; 2.5 mm, 3.96 [2.24 to 6.45]; p= 0.358), or NBD (median [range]: 5 mm, 7.5 Gy [2.5 to 12.4]; 2.5 mm, 7.5 Gy [2.5 to 12.5]; p= 0.675). The performance of the 5-mm MLC was not inferior to the 2.5-mm MLC in applications of noncoplanar VMAT-SRT for multiple brain metastases with regards to dose conformity, gradient, and NBD. This study provides the necessary background for generalizing noncoplanar VMAT-SRT approaches in treating multiple brain lesions.
AB - The aim of this study was to examine the feasibility of noncoplanar volumetric modulated arc stereotactic radiotherapy (VMAT-SRT) using a 5-mm multileaf collimator (MLC) for multiple brain metastases. We identified 34 multiple-target cases (3 to 19 targets in each case) with a total of 257 of targets and constructed noncoplanar VMAT-SRT plans using 5-mm and 2.5-mm MLCs with 4-arc. The prescribed dose was 36 Gy/6 fr. Plans were evaluated using the Paddick conformity indices (PCI), Paddick gradient index (PGI), and normal brain dose (NBD, equal to the mean brain dose minus gross tumor volume). There were no significant differences in PCI (median [range]: 5 mm, 0.88 [0.78 to 0.94]; 2.5 mm, 0.89 [0.78 to 0.94]; p= 0.691), PGI (median [range]: 5 mm, 3.96 [2.21 to 6.63]; 2.5 mm, 3.96 [2.24 to 6.45]; p= 0.358), or NBD (median [range]: 5 mm, 7.5 Gy [2.5 to 12.4]; 2.5 mm, 7.5 Gy [2.5 to 12.5]; p= 0.675). The performance of the 5-mm MLC was not inferior to the 2.5-mm MLC in applications of noncoplanar VMAT-SRT for multiple brain metastases with regards to dose conformity, gradient, and NBD. This study provides the necessary background for generalizing noncoplanar VMAT-SRT approaches in treating multiple brain lesions.
KW - Brain metastases
KW - Leaf width
KW - Noncoplanar arc
KW - Stereotactic radiotherapy
KW - Volumetric modulated arc therapy
UR - http://www.scopus.com/inward/record.url?scp=85071112252&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85071112252&partnerID=8YFLogxK
U2 - 10.1016/j.meddos.2019.07.006
DO - 10.1016/j.meddos.2019.07.006
M3 - Article
C2 - 31466736
AN - SCOPUS:85071112252
SN - 0958-3947
VL - 45
SP - 97
EP - 101
JO - Medical Dosimetry
JF - Medical Dosimetry
IS - 1
ER -