Feasibility of 5-mm vs 2.5-mm width multileaf collimator in noncoplanar volumetric modulated arc stereotactic radiotherapy for multiple brain metastases

Kotaro Yoshio, Akihisa Wakita, Kento Hisazumi, Takahiro Kitayama, Nobuhisa Tajiri, Tsuyoki Shiode, Shiro Akaki, Susumu Kanazawa

Research output: Contribution to journalArticlepeer-review

3 Citations (Scopus)

Abstract

The aim of this study was to examine the feasibility of noncoplanar volumetric modulated arc stereotactic radiotherapy (VMAT-SRT) using a 5-mm multileaf collimator (MLC) for multiple brain metastases. We identified 34 multiple-target cases (3 to 19 targets in each case) with a total of 257 of targets and constructed noncoplanar VMAT-SRT plans using 5-mm and 2.5-mm MLCs with 4-arc. The prescribed dose was 36 Gy/6 fr. Plans were evaluated using the Paddick conformity indices (PCI), Paddick gradient index (PGI), and normal brain dose (NBD, equal to the mean brain dose minus gross tumor volume). There were no significant differences in PCI (median [range]: 5 mm, 0.88 [0.78 to 0.94]; 2.5 mm, 0.89 [0.78 to 0.94]; p= 0.691), PGI (median [range]: 5 mm, 3.96 [2.21 to 6.63]; 2.5 mm, 3.96 [2.24 to 6.45]; p= 0.358), or NBD (median [range]: 5 mm, 7.5 Gy [2.5 to 12.4]; 2.5 mm, 7.5 Gy [2.5 to 12.5]; p= 0.675). The performance of the 5-mm MLC was not inferior to the 2.5-mm MLC in applications of noncoplanar VMAT-SRT for multiple brain metastases with regards to dose conformity, gradient, and NBD. This study provides the necessary background for generalizing noncoplanar VMAT-SRT approaches in treating multiple brain lesions.

Original languageEnglish
Pages (from-to)97-101
Number of pages5
JournalMedical Dosimetry
Volume45
Issue number1
DOIs
Publication statusPublished - Mar 1 2020

Keywords

  • Brain metastases
  • Leaf width
  • Noncoplanar arc
  • Stereotactic radiotherapy
  • Volumetric modulated arc therapy

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Oncology
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Feasibility of 5-mm vs 2.5-mm width multileaf collimator in noncoplanar volumetric modulated arc stereotactic radiotherapy for multiple brain metastases'. Together they form a unique fingerprint.

Cite this