Head multidetector-row CT angiography with low kV settings: Improved vessel enhancement and noise reduction in phantom studies

Hiroyuki Hamaguchi, Takuya Akagawa, Youji Kimura, Koutarou Hashimoto, Kazutoshi Tsunou, Sachiko Goto, Yoshiharu Azuma

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

In order to improve the measurement accuracy of blood vessels, it was investigated how much contrast was improved and how much noise was increased, with low kV setting for Head multidetector-row computed tomography (CT, Aquilion64, Toshiba Medical Systems) angiography in phantom study. That is, using low (80or 100) kV instead of 120-140 kV should shift the mean photon energy closer to the K-edge of iodine and therefore improve the contrast between vessels filled with contrast medium and their surroundings. For the vascular models, we used urethane tube (47 Hounsfield unit (HU)) with two inner diameters (2and 4 mm) filled with contrast medium diluted to density (373 ± 14 HU). The urethane tube was fixed in a water-filled columnar acrylic container with a diameter of 20 cm. A Teflon resin pipe was inserted inside the acrylic container as a skull and two Teflon resin pipes with a diameter of 5 cm were placed near the center of the acrylic container as basis cranii. Then, the profile curve of the cross section of the vascular models, and physical image qualities characteristics such as modulation transfer function, S.D. and Wiener spectrum, were evaluated changing CT scan conditions. Furthermore, reducing the noise was tried, using mAs change, reconstruction functions or Quantum Denoising Filters (Toshiba Medical Systems). The maximum CT values of the profile of 80 kV increased from 1.5 to 1.7 times than that of 120 kV. The noise has increased evenly in all frequency band regions at the same time though the contrast rose in 2 and 4 mm vascular models. Quantum Denoising Filters were effective in noise reduction. Since there are the useful noise reduction methods, it can be said that vessel enhancement with low kV settings is realizable.

Original languageEnglish
Title of host publicationIFMBE Proceedings
EditorsSun I. Kim, Tae Suk Suh
PublisherSpringer Verlag
Pages1554-1558
Number of pages5
Edition1
ISBN (Print)9783540368397
DOIs
Publication statusPublished - 2007
Event10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of
Duration: Aug 27 2006Sept 1 2006

Publication series

NameIFMBE Proceedings
Number1
Volume14
ISSN (Print)1680-0737
ISSN (Electronic)1433-9277

Other

Other10th World Congress on Medical Physics and Biomedical Engineering, WC 2006
Country/TerritoryKorea, Republic of
CitySeoul
Period8/27/069/1/06

Keywords

  • Blood vessel enhancement
  • Computed tomography angiography
  • Low kV setting
  • Noise reduction
  • Phantom studies

ASJC Scopus subject areas

  • Bioengineering
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Head multidetector-row CT angiography with low kV settings: Improved vessel enhancement and noise reduction in phantom studies'. Together they form a unique fingerprint.

Cite this