TY - JOUR
T1 - Mean lung pressure during adult high-frequency oscillatory ventilation
T2 - An experimental study using a lung model
AU - Hirayama, Takahiro
AU - Nagano, Osamu
AU - Shiba, Naoki
AU - Yumoto, Tetsuya
AU - Sato, Keiji
AU - Terado, Michihisa
AU - Ugawa, Toyomu
AU - Ichiba, Shingo
AU - Ujike, Yoshihito
N1 - Publisher Copyright:
© 2014 by Okayama University Medical School.
PY - 2014
Y1 - 2014
N2 - In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.
AB - In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3ml/cmH2O) with or without a resistor (20cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30cmH2O;frequency, 5-15Hz (every 1Hz);airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.
KW - HFOV
KW - Mean airway pressure
KW - Mean lung pressure
UR - http://www.scopus.com/inward/record.url?scp=84920986023&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84920986023&partnerID=8YFLogxK
M3 - Article
C2 - 25519026
AN - SCOPUS:84920986023
SN - 0386-300X
VL - 68
SP - 323
EP - 329
JO - Acta medica Okayama
JF - Acta medica Okayama
IS - 6
ER -