Suture retention strength of expanded polytetraf luoroethylene (ePTFE) graft

Yoshinari Mine, Hideya Mitsui, Yu Oshima, Yasuharu Noishiki, Mikizo Nakai, Shunji Sano

Research output: Contribution to journalArticlepeer-review

23 Citations (Scopus)

Abstract

Our meticulous investigation of ePTFE graft breakage when a wire placed at the edge of an ePTFE graft was pulled, revealed that, depending on the breakage pattern, a break starts much earlier than the peak suture retention strength, which is the current international indicator for anastomotic-site break strength. Furthermore, the breakage patterns differ based on the thickness of the wire and the fiber direction of the ePTFE graft. Based on these findings, we advocate measuring the peak suture retention strength using 0.10-mm sutures and a standardized wire thickness in order to assess the anas-tomotic retention strength of ePTFE grafts.

Original languageEnglish
Pages (from-to)121-128
Number of pages8
JournalActa medica Okayama
Volume64
Issue number2
Publication statusPublished - Apr 2010
Externally publishedYes

Keywords

  • Anastomotic strength
  • Eptfe
  • Suture retention strength

ASJC Scopus subject areas

  • Biochemistry, Genetics and Molecular Biology(all)

Fingerprint

Dive into the research topics of 'Suture retention strength of expanded polytetraf luoroethylene (ePTFE) graft'. Together they form a unique fingerprint.

Cite this