TY - JOUR
T1 - Validation of pencil beam scanning proton therapy with multi-leaf collimator calculated by a commercial Monte Carlo dose engine
AU - Tominaga, Yuki
AU - Sakurai, Yusuke
AU - Miyata, Junya
AU - Harada, Shuichi
AU - Akagi, Takashi
AU - Oita, Masataka
N1 - Publisher Copyright:
© 2022 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, LLC on behalf of The American Association of Physicists in Medicine.
PY - 2022
Y1 - 2022
N2 - This study aimed to evaluate the clinical beam commissioning results and lateral penumbra characteristics of our new pencil beam scanning (PBS) proton therapy using a multi-leaf collimator (MLC) calculated by use of a commercial Monte Carlo dose engine. Eighteen collimated uniform dose plans for cubic targets were optimized by the RayStation 9A treatment planning system (TPS), varying scan area, modulation widths, measurement depths, and collimator angles. To test the patient-specific measurements, we also created and verified five clinically realistic PBS plans with the MLC, such as the liver, prostate, base-of-skull, C-shape, and head-and-neck. The verification measurements consist of the depth dose (DD), lateral profile (LP), and absolute dose (AD). We compared the LPs and ADs between the calculation and measurements. For the cubic plans, the gamma index pass rates (γ-passing) were on average 96.5% ± 4.0% at 3%/3 mm for the DD and 95.2% ± 7.6% at 2%/2 mm for the LP. In several LP measurements less than 75 mm depths, the γ-passing deteriorated (increased the measured doses) by less than 90% with the scattering such as the MLC edge and range shifter. The deteriorated γ-passing was satisfied by more than 90% at 2%/2 mm using uncollimated beams instead of collimated beams except for three planes. The AD differences and the lateral penumbra width (80%–20% distance) were within ±1.9% and ± 1.1 mm, respectively. For the clinical plan measurements, the γ-passing of LP at 2%/2 mm and the AD differences were 97.7% ± 4.2% on average and within ±1.8%, respectively. The measurements were in good agreement with the calculations of both the cubic and clinical plans inserted in the MLC except for LPs less than 75 mm regions of some cubic and clinical plans. The calculation errors in collimated beams can be mitigated by substituting uncollimated beams.
AB - This study aimed to evaluate the clinical beam commissioning results and lateral penumbra characteristics of our new pencil beam scanning (PBS) proton therapy using a multi-leaf collimator (MLC) calculated by use of a commercial Monte Carlo dose engine. Eighteen collimated uniform dose plans for cubic targets were optimized by the RayStation 9A treatment planning system (TPS), varying scan area, modulation widths, measurement depths, and collimator angles. To test the patient-specific measurements, we also created and verified five clinically realistic PBS plans with the MLC, such as the liver, prostate, base-of-skull, C-shape, and head-and-neck. The verification measurements consist of the depth dose (DD), lateral profile (LP), and absolute dose (AD). We compared the LPs and ADs between the calculation and measurements. For the cubic plans, the gamma index pass rates (γ-passing) were on average 96.5% ± 4.0% at 3%/3 mm for the DD and 95.2% ± 7.6% at 2%/2 mm for the LP. In several LP measurements less than 75 mm depths, the γ-passing deteriorated (increased the measured doses) by less than 90% with the scattering such as the MLC edge and range shifter. The deteriorated γ-passing was satisfied by more than 90% at 2%/2 mm using uncollimated beams instead of collimated beams except for three planes. The AD differences and the lateral penumbra width (80%–20% distance) were within ±1.9% and ± 1.1 mm, respectively. For the clinical plan measurements, the γ-passing of LP at 2%/2 mm and the AD differences were 97.7% ± 4.2% on average and within ±1.8%, respectively. The measurements were in good agreement with the calculations of both the cubic and clinical plans inserted in the MLC except for LPs less than 75 mm regions of some cubic and clinical plans. The calculation errors in collimated beams can be mitigated by substituting uncollimated beams.
KW - commissioning
KW - lateral penumbra
KW - multi-leaf collimator
KW - pencil beam scanning
KW - proton therapy
UR - http://www.scopus.com/inward/record.url?scp=85142535379&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85142535379&partnerID=8YFLogxK
U2 - 10.1002/acm2.13817
DO - 10.1002/acm2.13817
M3 - Article
C2 - 36420959
AN - SCOPUS:85142535379
SN - 1526-9914
JO - Journal of Applied Clinical Medical Physics
JF - Journal of Applied Clinical Medical Physics
ER -