Coil Occlusion for Patent Ductus Arteriosus in Japan

Hideshi Tomita, Shigeto Fuse, Teiji Akagi, Kazuyuki Koike, Masahiro Kamada, Tetsuro Kamiya, Kazuo Momma, Akira Ishizawa, Shunzo Chiba

研究成果査読

5 被引用数 (Scopus)

抄録

We surveyed Japanese experience of coil occlusion of patent ductus arteriosus up to 30 September 1996 by sending questionnaires to 175 hospitals. Thirty-four hospitals reported outcome data for 231 procedures in 218 patients. Successful implantation was achieved in 94% and acute complete closure of the ductus occurred in 71% of those in whom implantation of the coil was successful. Of the latter, 83% reported late complete closure. When those patients who underwent reocclusion for residual shunt are included, 89% attained complete closure. No life-threatening complications have occurred so far. Late reopening was reported in 3 cases. Although the angiographic type of ductus was significantly related to successful implantation (p<0.01), there was no significant correlation with complete occlusion. Ductuses with a minimum diameter greater than 3 mm had a decreased chance of successful implantation, whereas those less than 2 mm had a greater incidence of complete closure.

本文言語English
ページ(範囲)997-1003
ページ数7
ジャーナルJAPANESE CIRCULATION JOURNAL
61
12
DOI
出版ステータスPublished - 1997
外部発表はい

ASJC Scopus subject areas

  • 生理学
  • 循環器および心血管医学

フィンガープリント

「Coil Occlusion for Patent Ductus Arteriosus in Japan」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル