TY - JOUR
T1 - Compact and easy-operation magnetocardiograph with four-channel planar gradiometers
AU - Yokosawa, K.
AU - Suzuki, D.
AU - Tsukamoto, A.
AU - Miyashita, T.
AU - Kandori, A.
AU - Tsukada, K.
AU - Takagi, K.
PY - 2001/12/1
Y1 - 2001/12/1
N2 - A magnetocardiograph (MCG), which consists of a sensor comprising four superconducting-quantum-interference device (SQUID) gradiometers and a magnetic shielding cylinder made of nanocrystalline soft magnetic materials, has been developed. The sensor can be handled easily because the gradiometers are made of high-critical-temperature (Tc) superconductor operated in liquid nitrogen. Further, the shielding cylinder is lightweight (160 kg) and compact (2 m long and 1 m in diameter). The gradiometer balance is high enough (typically 0.1%) for recording magnetocardiograms inside the shielding cylinder, whose shielding factor is -35 dB at 1 Hz. We used the new MCG to record magnetocardiograms of a healthy volunteer at four different positions. From this magnetocardiograms we then obtained a current arrow map, by which mycardium activity can be estimated, at 16 sites (4 × 4 matrix) on the measurement plane. The similarity between the current-arrow map obtained by a conventional MCG and that from the newly developed MCG indicates that the developed compact MCG is also capable of estimating the region of cardiac activity.
AB - A magnetocardiograph (MCG), which consists of a sensor comprising four superconducting-quantum-interference device (SQUID) gradiometers and a magnetic shielding cylinder made of nanocrystalline soft magnetic materials, has been developed. The sensor can be handled easily because the gradiometers are made of high-critical-temperature (Tc) superconductor operated in liquid nitrogen. Further, the shielding cylinder is lightweight (160 kg) and compact (2 m long and 1 m in diameter). The gradiometer balance is high enough (typically 0.1%) for recording magnetocardiograms inside the shielding cylinder, whose shielding factor is -35 dB at 1 Hz. We used the new MCG to record magnetocardiograms of a healthy volunteer at four different positions. From this magnetocardiograms we then obtained a current arrow map, by which mycardium activity can be estimated, at 16 sites (4 × 4 matrix) on the measurement plane. The similarity between the current-arrow map obtained by a conventional MCG and that from the newly developed MCG indicates that the developed compact MCG is also capable of estimating the region of cardiac activity.
KW - High-Tc
KW - Magnetocardiograph
KW - Shielding
KW - SQUID
UR - http://www.scopus.com/inward/record.url?scp=0035781455&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0035781455&partnerID=8YFLogxK
M3 - Conference article
AN - SCOPUS:0035781455
SN - 0454-9244
VL - 4
SP - 3281
EP - 3284
JO - Annual Reports of the Research Reactor Institute, Kyoto University
JF - Annual Reports of the Research Reactor Institute, Kyoto University
T2 - 23rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society
Y2 - 25 October 2001 through 28 October 2001
ER -