Comparison of the ventilation characteristics in two adult oscillators: a lung model study

Tetsuya Yumoto, Takahisa Fujita, Sunao Asaba, Shunsuke Kanazawa, Atsunori Nishimatsu, Hideo Yamanouchi, Satoshi Nakagawa, Osamu Nagano

研究成果査読

1 被引用数 (Scopus)

抄録

Background: Two recent large randomized controlled trials did not show the superiority of high-frequency oscillatory ventilation (HFOV) in adults with ARDS. These two trials had differing results, and possible causes could be the different oscillators used and their different settings, including inspiratory time % (IT%). The aims of this study were to obtain basic data about the ventilation characteristics in two adult oscillators and to elucidate the effect of the oscillator and IT% on ventilation efficiency. Methods: The Metran R100 or SensorMedics 3100B was connected to an original lung model internally equipped with a simulated bronchial tree. The actual stroke volume (aSV) was measured with a flow sensor placed at the Y-piece. Carbon dioxide (CO2) was continuously insufflated into the lung model (V ̇ CO2), and the partial pressure of CO2 (PCO2) in the lung model was monitored. Alveolar ventilation (V ̇ A; L/min) was estimated as V ̇ CO2 divided by the stabilized value of PCO2. V ̇ A was evaluated with several stroke volume settings in the R100 (IT = 50%) or several airway pressure amplitude settings in the 3100B (IT = 33%, 50%) at a frequency of 6 and 8 Hz, a mean airway pressure of 25 cmH2O, and a bias flow of 30 L/min. Assuming that V ̇ A = frequencya × aSVb, values of a and b were determined. Ventilation efficiency was calculated as V ̇ A divided by actual minute ventilation. Results: The relationship between aSV and V ̇ A or ventilation efficiency were different depending on the oscillator and IT%. The values of a and b were 0 < a < 1 and 1 < b < 2 and were different for three conditions at both frequencies. V ̇ A and ventilation efficiency were highest with R100 (IT = 50%) and lowest with 3100B (IT = 33%) for high aSV ranges at both frequencies. Conclusions: In this lung model study, ventilation characteristics were different depending on the oscillator and IT%. Ventilation efficiency was highest with R100 (IT = 50%) and lowest with 3100B (IT = 33%) for high aSV ranges.

本文言語English
論文番号15
ジャーナルIntensive Care Medicine Experimental
7
1
DOI
出版ステータスPublished - 12月 1 2019
外部発表はい

ASJC Scopus subject areas

  • 集中医療医学

フィンガープリント

「Comparison of the ventilation characteristics in two adult oscillators: a lung model study」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル