抄録
In fabrication of a compact NMR (Nuclear Magnetic Resonance) magnet which consists of a stacked HTS (High Temperature Superconducting) bulk annuli, generally there are three key issues: spatial homogeneity, temporal stability, strength of trapped magnetic fields. This paper presents a study on the effects of axial gap length between stacked HTS bulks on the three key issues of a bulk HTS magnet for compact HTS NMR applications. The HTS bulk magnet of which the ID and OD are 20 and 60 mm respectively has a 50 and 80 mm heights depending on the axial gap lengths between HTS bulks. The gap length between each HTS bulk varied from 0 mm to 10 mm and were used as parameters to optimize, analytically as well as experimentally, the overall field homogeneity of the HTS bulk magnet. The optimized axial gap length was obtained by analytical results, and the better magnetic field homogeneity and temporal stability of trapped magnetic field were achieved by lower magnetization field. The improved spatial homogeneity and strength of generated magnetic field by a new compact NMR magnet will be presented.
本文言語 | English |
---|---|
論文番号 | 6119203 |
ジャーナル | IEEE Transactions on Applied Superconductivity |
巻 | 22 |
号 | 3 |
DOI | |
出版ステータス | Published - 2012 |
ASJC Scopus subject areas
- 電子材料、光学材料、および磁性材料
- 凝縮系物理学
- 電子工学および電気工学