N2 - To establish the electrophysiological mappings of nonhuman primates by using magnetocardiogram (MCG) data and obtain the normal values of MCG parameters, we used 64-channel superconducting quantum interference devices to measure 8 × 8 MCG data for 95 cynomolgus monkeys (Macaca fascicularis, 51 female and 44 male). The PQ interval, QRS duration, QT interval and QTc were respectively 79 ± 14 ms, 42 ± 7 ms, 222 ± 23 ms and 363 ± 25 (mean ± SD), and these parameters did not differ significantly between female and male monkeys. These results indicate the normal values of the MCG parameters of the cynomolgus monkey and should facilitate animal experiments in magnetocardiography.
AB - To establish the electrophysiological mappings of nonhuman primates by using magnetocardiogram (MCG) data and obtain the normal values of MCG parameters, we used 64-channel superconducting quantum interference devices to measure 8 × 8 MCG data for 95 cynomolgus monkeys (Macaca fascicularis, 51 female and 44 male). The PQ interval, QRS duration, QT interval and QTc were respectively 79 ± 14 ms, 42 ± 7 ms, 222 ± 23 ms and 363 ± 25 (mean ± SD), and these parameters did not differ significantly between female and male monkeys. These results indicate the normal values of the MCG parameters of the cynomolgus monkey and should facilitate animal experiments in magnetocardiography.
UR - http://www.scopus.com/inward/record.url?scp=41049086929&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=41049086929&partnerID=8YFLogxK
U2 - 10.1088/0031-9155/53/6/007
DO - 10.1088/0031-9155/53/6/007
M3 - Article
C2 - 18367791
AN - SCOPUS:41049086929
SN - 0031-9155
VL - 53
SP - 1609
EP - 1618
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 6
ER -